87
Overpressure Protector Adult/Pediatric: 300 mmHg (N.C), 330 mmHg (S.F.C)
Neonatal: 150 mmHg (N.C), 165 mmHg (S.F.C)
Defibrillator Protection Protected
Measurement Speed About 20 seconds
At the following condition:
Adult
Cuff size 12 cm
SYS 120 mmHg
MAP 90 mmHg
DIA 80 mmHg/ PR 80 BPM
Manual Measurements (180 mmHg)
SpO
2
Pulse Rate
Range Nellcor module: 20 to 300 BPM
Mediana module: 30 to 300 BPM
Accuracy Nellcor module: ±3 BPM
Mediana module: ±2 % or 2 BPM, whichever is greater
SpO
2
Range Nellcor module: 1 to 100 %
Mediana module:
0 to 99 %
Low Perfusion
(Nellcor module only)
0.03 to 20 %
Accuracy Nellcor module:
Without Interference-Adult/Pediatric
70 to 100 % ±2 digits
1 to 69 % unspecified
Without Interference-Neonatal
70 to 100 % ±3 digits
1 to 69 % unspecified
Low Perfusion 70 to 100 % ±2 digits
1 to 69 % unspecified
Mediana module:
Without Interference-Adult/Pediatric/Neonatal
70 to 99 % ±2 digits
0 to 69 % unspecified
Display Update Within 30 seconds
Display Sweep Speeds 12.5 mm/sec, 25.0 mm/sec and 50.0 mm/sec
Defibrillator Protection Protected
Note: Neonatal specifications are shown for neonatal sensors with the monitor. Saturation
accuracy will vary by sensor type as specified by the manufacturer.
Note: The wavelength range of the light emitted are 660nm and 890nm with the energy not
exceeding 15mW (for Nellcor module) and 660nm with the energy not exceeding 2mW
and 905nm with the energy not exceeding 2~2.4mW (for Mediana module).