68
Figure 22. Rear Case Disassembly - Main Board, Side Connector Cover, ECG
Connector, CO
2
Connector
Table 23. Part Descriptions - Main Board, Side Connector Cover, ECG Connector, CO
2
Connector
Part Codes Descriptions Qty
P1105 / P1093 Main Board for Nellcor SpO
2
module / for Mediana SpO
2
module 1/1
T0212 Side connector cover 1
E4114 Luer fitting/NIBP connector 1
E4231 ECG connector (6pin) 1
W0158 EtCO
2
connector (8pin) 1
C4. Side connector cover disassembly
1. Remove the side connector from the inner case.
2. Remove the luer fitting/NIBP connector from the side connector cover.
C4-1. ECG connector disassembly
1.
Remove 4 tapping screws (3x8) from the ECG connector.
2.
Remove the ECG connector from the side connector cover.
C5. Main Board disassembly
1. Remove 5 round-head screws (3x6) from the Main Board.
2. Separate the ECG arrester cable, DC power cable, TCP/IP cable, SMPS control cable,
external DC cable and battery cable from the main board.
3. Remove the Main Board from the inner case assembly.
Kommentare zu diesen Handbüchern