Mediana P10 Wartungshandbuch Seite 38

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 110
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 37
32
SAFETY TEST
TEST CONDITIONS LIMIT (uA) RESULTS REMARKS
Earth leakage current (NC) 500 Pass / Fail Value: uA
Earth leakage current (SFC OS) 1000 Pass / Fail Value: uA
Earth leakage current (NCRM) 500 Pass / Fail Value: uA
Earth leakage current (SFC OSRM) 1000 Pass / Fail Value: uA
Enclosure leakage current (NC) 100 Pass / Fail Value: uA
Enclosure leakage current (OS) 500 Pass / Fail Value: uA
Enclosure leakage current (SFC OE) 500 Pass / Fail Value: uA
Enclosure leakage current (NCRM) 100 Pass / Fail Value: uA
Enclosure leakage current (SFC OSRM) 500 Pass / Fail Value: uA
Enclosure leakage current (SFC OERM) 500 Pass / Fail Value: uA
Patient leakage current (NC) 10 Pass / Fail Value: uA
Patient leakage current (OS) 50 Pass / Fail Value: uA
Patient leakage current (SFC OE) 50 Pass / Fail Value: uA
Patient leakage current (NCRM) 10 Pass / Fail Value: uA
Patient leakage current (SFC OSRM) 50 Pass / Fail Value: uA
Patient leakage current (SFC OERM) 50 Pass / Fail Value: uA
Mains voltage on applied part (SFC) 50 Pass / Fail Value: uA
Mains voltage on applied part (SFCRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG LL (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG LL (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG LL (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG LL (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG LL (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG LL (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG RA (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG RA (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG RA (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG RA (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG RA (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-ECG RA (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-ECG RA (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL-ECG RA (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-ECG RA (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-ECG RA (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL-ECG RA (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-ECG RA (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO
2
(NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO
2
(OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO
2
(SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO
2
(NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO
2
(SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO
2
(SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO
2
(NC) 10 Pass / Fail Value: uA
Seitenansicht 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 109 110

Kommentare zu diesen Handbüchern

Keine Kommentare