PERFORMANCE VERIFICATION
General
This section discusses the tests used to verify performance following repairs or during
routine maintenance. All tests can be performed without removing the monitor covers. All
tests except the battery charge and battery discharge tests must be performed as the last
operation before the monitor is returned to the user.
If the monitor fails to perform as specified in any test, repairs must be made to correct the
problem before the monitor is returned to the user.
Required Equipment
Table 1 lists the equipment required for performance verifications.
Table 1. Required Equipment
Equipment Description
Digital multimeter (DMM) Fluke Model 87 or equivalent
ECG cable for 3 leads ECG trunk cable for 3 leads
ECG cable for 5 leads (option) ECG trunk cable for 5 leads
ECG 3 lead wires ECG 3 leads wire pack (SNAP)
ECG 5 lead wires (option) ECG 5 leads wire pack (SNAP)
NIBP cuff hose Omron cuff hose No.1
NIBP cuff Omron cuff HEM series
Dummy can Dummy can-large
Dummy can Dummy can-small
SpO
2
extension cable Nellcor DOC-10 or RCP058
SpO
2
sensor (durable) Nellcor DS-100A or YM-1
Temperature probes YSI-400 series
Mainstream CO
2
sensor (option) Capnostat 5 mainstream CO
2
sensor
Mainstream CO
2
monitoring
airway adapter (option)
TruLink adult reusable airway adapter
Sidestream CO
2
sensor (option) LoFlo sidestream CO
2
sensor
Sidestream CO
2
airway adapter
and sampling line (option)
TruLink adult/pediatric airway adapter and
sampling line
ECG simulator Metron PS-420 or equivalent
SpO
2
simulator Nellcor SRC-MAX simulator (for Nellcor module)
SpO
2
simulator ITEC Engineering LLC 8010 oximetry simulator
(for Mediana module)
NIBP simulator Bio-Tek BP Pump 2 or equivalent
Temperature simulator medSim 300 or equivalent
Respiration simulator Metron PS-420 or equivalent
CO
2
gas flow meter STEC SEF-21A or equivalent
CO
2
gas cylinder 10% CO
2
gas cylinder
Safety analyzer Metron QA-90 or equivalent
Stopwatch Manual or electronic
Note: The sphygmomanometer must be calibrated periodically. The correct value can not
be found if the sphygmomanometer has not been calibrated.
Note: Contact Mediana Technical Service Department for pricing and ordering information
of the required equipment.